您现在的位置是:网站首页>行业标准JJF(赣)020-2021 收敛计校准规范

JJF(赣)020-2021 收敛计校准规范

简介
JJF(赣)020-2021
收敛计校准规范
Calibration Specification for Tape Extensometer
2021-11-09发布
2022-03-01 实施 目录 引言.................................................................(1) 1 范围...............................................................(1) 2 引用文件...........................................................(1) 3 概述...............................................................(1) 4 计量特性...........................................................(1) 4.1 张力.............................................................(1) 4.2 示值误差.........................................................(1) 4.3 漂移.............................................................(2) 4.4 测量重复性.......................................................(2) 5 校准条件...........................................................(2) 5.1 环境条件.........................................................(2) 5.2 测量标准及其他设备...............................................(2) 6 校准项目和校准方法.................................................(2) 6.1 校准项目.........................................................(2) 6.2 校准方法.........................................................(3) 7 校准结果表达.......................................................(5) 8 复校时间间隔.......................................................(6) 附录 A 收敛计精测装置的示值误差测量结果的不确定度评定..................(7) 1 范围 本规范适用于测量范围为(0.5 ~50)m 收敛计的校准。 2 引用文件 本规范引用下列文件: MT/T919-2002 电子数显式收敛计 凡是注日期的引用文件,仅注日期的版本适用于本规范;凡是不注日期的引用文件,其最新版本(包括所有的修改单)适用于本规范。 3 概述 收敛计是用于测量隧道、巷道、峒室及其它工程围岩周边任意方向两点间距离的微小变化的计量器具。利用机械传递位移的方法,将两个基准点间的相对位移转变为数显位移计的两次读数差。仪器典型结构如图 1 所示。